Skip to main navigation Skip to search Skip to main content

Application of hydride generation to the determination of trace concentrations of arsenic by capacitively coupled microwave plasma

Research output: Contribution to journal › Article › peer-review

Abstract

Studies are presented describing an application of the NaBH4 reduction of soluble arsenite to form arsine for ultra-trace level arsenic determination by capacitively coupled microwave plasma atomic emission spectrometry. The arsine was condensed in a liquid nitrogen cooled trap, and introduced into the plasma after the elimination of reaction by-products that would reduce the excitation efficiency of the plasma. Use of a photodiode array detector provided a means of background correction. The procedure resulted in a limit of detection of 110 pg and linear dynamic range of 104 for arsenic.
Original languageEnglish
Pages (from-to)279-285
Number of pages7
JournalAnalytica Chimica Acta
Volume278
Issue number2
DOIs
Publication statusPublished - 1993

Fingerprint

Dive into the research topics of 'Application of hydride generation to the determination of trace concentrations of arsenic by capacitively coupled microwave plasma'. Together they form a unique fingerprint.