Abstract
Three-dimensional imaging with 2-D matrix probes is one of the most exciting recent ultrasound innovations. Unfortunately, the number of elements of a 2-D matrix probe is often very high, and reducing this number deteriorates the beam properties. In this paper, we propose a new sparse-array design technique with irregular element positioning, which significantly reduces the number of active elements as well as the grating-lobe level. In particular, we introduce a new cost function for optimizing the weighting coefficients of the elements and a new annealing-based algorithm to compute the lowest cost solutions. Numerical simulations show substantial improvements over standard sparse arrays.
| Original language | English |
|---|---|
| Article number | 6529099 |
| Pages (from-to) | 3093-3102 |
| Number of pages | 10 |
| Journal | IEEE Transactions on Biomedical Engineering |
| Volume | 60 |
| Issue number | 11 |
| DOIs | |